Measurement technology for vacuum insulation. What happens inside a vacuum body cannot be seen from the outside

Jochen Hiemeyer
Head of Technology
Technology
Category
November 2026
Published

Why good vacuum insulation starts with good measurement technology and how we measure at TerraVac.

The short answer

Whether vacuum insulation lives up to its promise over the years is determined by pressures and gas volumes that you can neither see nor feel. Depending on the core material, even one millibar of gas pressure or a fraction of a milligram of water can significantly impair the insulation effect. That is why we measure not only the material, but also its behaviour in a vacuum and the finished product. We have developed many of the measurement setups for this ourselves.

Key takeaways

  • Vacuum insulation loses its effectiveness long before anything is visible on the outside.
  • Heat is harder to measure than electricity because there is no such thing as a perfect thermal insulator.
  • We measure thermally on three levels: the thermal conductivity of the material, a rapid dynamic test, and the heat loss of the overall product.
  • In addition, there is permeation, internal pressure and gas composition, leak detection, flow, sorption, and mechanics.
  • For you, this means: statements about the long-term behaviour of your product are based on measurement data, not assumptions.

Why declining vacuum often remains invisible

When people think of defective vacuum insulation, most think of a hole. Air flows in, the casing becomes soft, and the body loses its shape. That is the obvious case. More critical are changes that are much smaller and not noticeable from the outside.

‍

Minimal gas pressure, major impact. If the gas pressure inside vacuum insulation with a glass fibre core rises to just 1 mbar, thermal conductivity will have already more than quintupled. Mechanically, this goes entirely unnoticed. The pressure difference is smaller than natural fluctuations in atmospheric pressure. Simply carrying a panel ten metres higher results in an air pressure change of more than one millibar.

‍

Minute amounts of water. In a panel with a void volume of around one litre, less than a single milligram of unbonded internal water can trigger a similarly sharp increase in thermal conductivity.

‍

Water turns into hydrogen. Ingressing water can react with metals to produce hydrogen—a gas that conducts heat far more effectively than air. In this way, even minor moisture ingress escalates into a significant thermal issue.

‍

Whether vacuum insulation will last for the service life of your product cannot be judged by looking at it. You have to measure it, and with such precision that even these small effects become visible.

Why heat is so difficult to measure

In electrical engineering, measuring is relatively simple. There are around 25 orders of magnitude between the best electrical conductor and the best insulator. Current flows where it is supposed to, and nowhere else.

‍

With heat, there are only about six orders of magnitude, from the best thermal conductor, diamond, to the best insulating material, vacuum insulation. There is no such thing as a perfect thermal insulator. Every mount, every sensor, and every cable conducts heat as well. With vacuum insulation, the heat flow you want to measure is so small that such side paths quickly become larger than the actual signal.

‍

Anyone who wants to develop vacuum insulation and manufacture it in series therefore needs measurements whose errors are smaller than the effects they are looking for. This is the standard to which we align our measurement technology.

Level 1: Precisely determining thermal conductivity

For the direct measurement of thermal conductivity, we use an external measuring device. It is one of the few fields in which we do not rely on our own measurement setups. We searched for and found a specialist for this. We work with the λ-Meter EP500 from Lambda-Messtechnik GmbH in Dresden.

‍

A vacuum insulation sample is placed in the λ-Meter EP500.

‍

The device operates using the guarded hot plate method (in accordance with ISO 8302 and DIN EN 12667). The sample is placed between a heated plate and a cooled plate. Measurements are taken only in the centre of the sample. A heated and a cooled guard ring around it ensure that no heat escapes laterally. Only when a stable state has been reached is the thermal conductivity determined from the power input, the temperature difference, the sample thickness, and the measuring area. This takes hours per measurement point, but provides an absolute value that does not depend on a calibration sample. This measuring instrument convinced us, especially with the very small heat flows that occur with vacuum insulation.

‍

However, a convincing measurement concept alone is not enough. Whether the implementation in the device really works only becomes apparent through series of measurements. The challenge here is to find samples that are uniform enough to be measured reproducibly over a long period of time.

‍

We deliberately made it difficult for ourselves. Since autumn 2023, we have been repeatedly measuring the heat transmission of three real vacuum freezer doors. Because a door is not a homogeneous material, we evaluate the heat transmission (U-value) rather than the thermal conductivity. For a plate device, this is the most difficult measurement situation imaginable. There is metal on the outside, uneven plastic on the inside, and 65 mm of STRATMO in between – so there is no clean plate contact. The doors protrude far beyond the measuring device and therefore couple with the fluctuating room temperature. And the measurements have been taken by different people over the years.

‍

Nevertheless, the values for each door have remained within a range of a few percent over almost three years, and the fluctuations are largely parallel. Door 1 and Door 3 are only about five percent apart and yet are clearly separated from each other at every point in time, always in the same order. For you, this means that a difference we measure between two variants is real and not measurement noise.

‍

Heat transmission (U-value) of three freezer doors, measured from October 2023 to August 2026.

Level 2: Quickly identifying when something is wrong

Highly accurate thermal conductivity measurement is only part of the picture. If you want to find leaks, you cannot wait hours for a steady state for every single part. This requires dynamic thermal measurement, which provides a sufficiently accurate estimate of thermal conductivity in a short time.

‍

To achieve this, we have developed our own heat flux sensor. It has a very low thermal mass and very low thermal resistance. This means it alters the heat flow it is intended to measure as little as possible. At the same time, it is designed for high measurement resolution and captures even very small changes in heat flow. The sensor records the actual progression rather than waiting for a stable final state. This allows us to identify problematic parts in a short measurement time, making the process suitable for series testing.

‍

Michael Freitag, head of our measurement laboratory, with the custom-developed heat flux sensor.

Level 3: Measuring the complete product

Ultimately, it is not the value on the data sheet that counts, but how much heat enters your product, including edges, joints, and feed-throughs. That is why we also measure the complete product.

‍

To do this, we apply a defined heating output to the insulation product and measure the temperature difference that develops relative to the environment. The quotient of the heating output and the temperature difference gives the heat loss value, the so-called U·A value. The principle is simple. The crucial factor is to record the ambient conditions precisely and determine the output exactly, as the heat flows here are also small.

‍

For you, this value is the most meaningful. It shows how the insulation performs in the system, not just what the material is capable of in the laboratory.

What we measure beyond thermal properties

Thermal measurement shows that something has changed. Other measurements show why. We have also developed our own measurement setups for this purpose.

  • Permeation: We determine how much gas enters the vacuum body through films or components, even at minimal rates. We are often more accurate in this than manufacturer specifications allow, which enables us to select casing materials based on our own data. Internal pressure and gas composition: In a destructive test, we create a sealed access point to the vacuum chamber to see which partial pressures prevail inside. This reveals whether air, water vapour, or hydrogen is the cause. Leak detection and characterisation: Helium leak detection, microscopic examinations, and accelerated ageing show where and how a vacuum body is leaking. Flow conductance: We measure and simulate how gases move in all flow ranges, from normal pressure to high vacuum. This is the basis for understanding evacuation and gas distribution in the core. Sorption and desorption: We characterise how gases accumulate on surfaces and are bound there physically or chemically, tailored to the conditions in the vacuum insulation. Mechanics: We measure and simulate flexural rigidity and compressive strength, particularly in combination with the cover layers.

What this means for your product

One of the most common questions about vacuum insulation is whether the vacuum remains stable over its service life. An honest answer to this can only be provided with measurement data. That is why, for us, measurement technology is not a testing phase at the end of development, but part of every stage. In the Proof phase, we measure your first prototypes; in the Shape phase, we validate the design; and in the Launch and Scale phases, we monitor the series.

‍

And because we measure everything from the material to the complete product, we view your application as a system. A material property alone does not tell you how a device, a storage unit, or a panel will insulate in use. The heat loss of the complete product does.

Frequently asked questions

Can you tell by looking at vacuum insulation whether it is intact?

Only in the case of a major defect. A pressure increase that already significantly impairs the insulation effect remains invisible from the outside. This requires thermal and vacuum-technical measurements.

‍

Why do you measure thermal conductivity with an external device?

There are specialised devices for stationary absolute measurement whose measurement principle we find convincing and whose results we have verified over long series of measurements. Our own expertise lies in the selection, sample preparation, evaluation, and all other measurements related to the vacuum.

‍

What is the difference between thermal conductivity and the U·A value?

Thermal conductivity describes the material. The U·A value describes how much heat flows through an entire product per degree of temperature difference, including all edges and penetrations. The latter value is the decisive one for your product design.

Conclusion

What happens inside a vacuum body cannot be seen. However, it can be measured, provided the measurement uncertainty is smaller than the effects being sought. We build on this with precise thermal analysis from material to product, and with our own test setups for what happens in a vacuum.

‍